OrbiSinus
Indications:
Spiral mode for orbital and paranasal sinuses studies,
e. g. sinusitis, mucocele, pneumatization, polyposis,
tumor, corrections etc.
Two recon jobs are predefined for image reconstruction
of the soft tissue and bone structure.
A range of 40 cm will be covered in 8.7 sec.
Head
OrbiSinus 2
nd
reconstr.
kV 130
Effective mAs 60
Slice collimation 1.0 mm
Slice width 3.0 mm 1.25 mm
Feed/Rotation 6.0 mm
Rotation time 1.0 sec.
Kernel H31s H70s
Increment 3.0 mm 0.8 mm
Image order ca-cr
CTDI
Vol
16.5 mGy
90

Table of Contents