36
Children
Head
Indications:
Spiral mode for routine head studies, e. g. tumors,
hydrocephalus, hemorrhaging, abnormalities, etc.
A range of 8.5 cm will be covered in 16.2 sec.
Head
kV 110
Effective mAs *
Slice collimation 1.0 mm
Slice width 5.0 mm
Feed/Rotation 6.0 mm
Rotation time 1.0 sec.
Kernel C30s
Increment 5.0 mm
Image order ca-cr
Tips
Children, who are more than 6 years old, should
be scanned with an adult protocol as the skull by this
time is fully grown.
When bone structure is of interest, use kernel C60s
for image reconstruction.
Posterior Fossa Optimization Filter
PFO: To reduce beam-hardening artifacts in head
images, particularly in the base of the skull, use the
Posterior Fossa Optimization (PFO) filter.
Contrast medium IV injection
Start delay exam dependent
Flow rate dependent upon needle size/Access site
Total amount 1 2 ml per kg of body weight
* The mAs should be adjusted to the age of the child.
Body Weight kV mAs CTDI
Vol
(mGy)
< 6 months 110 60 11.4
6 months 3 years 110 170 32.3
3 years 6 years 110 310 59.0