InnerEar
Indications:
Spiral mode for inner ear High Resolution studies,
e. g. inflammatory changes, tumorous processes
of pyramids, cerebellopontine angle tumors, post-
traumatic changes, etc.
A range of 40 cm will be covered in 15.3 sec.
Head
InnerEar
kV 130
Effective mAs 150
Slice collimation 0.5 mm
Slice width 0.63 mm
Feed/Rotation 3.0 mm
Rotation time 1.0 sec.
Kernel H90s
Increment 0.5 mm
Image order ca-cr
CTDI
Vol
42.7 mGy
For image reconstruction of soft tissue, use kernel
H30s/H31s.
Contrast medium IV injection
Start delay 60 sec.
Flow rate 2 ml/sec.
Total amount 50 ml
88