LungLowDose
kV 110
Effective mAs 16
Slice collimation 2.0 mm
Slice width 5.0 mm
Feed/Rotation 12.0 mm
Rotation time 0.8 sec.
Kernel B50s
Increment 5.0 mm
Image order cr-ca
CTDI
Vol
1.3 mGy
111
You could repeat the same protocol simply by clicking
the chronicle with the right mouse button for “repeat“.
E. g. when both non-contrast and contrast studies are
required.
For lung cancer staging, this protocol can be combined
with protocol “Neck Routine“.
Contrast medium IV injection
Start delay 30 sec.
Flow rate 2.5 ml/sec.
Total amount 50 70 ml