Thorax
ThoraxCombi 2
nd
reconstr.
kV 130
Effective mAs 70
Slice collimation 1.0 mm
Slice width 5.0 mm 1.25 mm
Feed/Rotation 10.8 mm
Rotation time 0.8 sec.
Kernel B41s B70s
Increment 5.0 mm 0.8 mm
Image order cr-ca
CTDI
Vol
9.2 mGy
If you want to reconstruct thin slices in every 2 mm
or 10 mm instead of 0.8 mm as predefined, simply
change the increment before image reconstruction.
Contrast medium IV injection
Start delay 25 sec.
Flow rate 2.5 ml/sec.
Total amount 80 ml
MIPthin: ThoraxCombi coronal
Image thickness 3
Distance between images 3
Number of images 60
For the 2
nd
reconstruction the Autoload into MIPthin
Range on the 3D Card is activated. The images will be
automatically loaded into 3D, MIPthin, and a coronal
MIPthin Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.