129
Thorax
ThoraxFast
kV 130
Effective mAs 70
Slice collimation 3.0 mm
Slice width 6.0 mm
Feed/Rotation 27.0 mm
Rotation time 0.6 sec.
Kernel B41s
Increment 6.0 mm
Image order cr-ca
CTDI
Vol
7.5 mGy
You could repeat the same protocol simply by clicking
the chronicle with the right mouse button for “repeat“.
E. g. when both non-contrast and contrast studies are
required.
For lung cancer staging, this protocol can be combined
with protocol “Neck Routine”.
Contrast medium IV injection
Start delay 25 30 sec.
Flow rate 2.5 ml/sec.
Total amount 80 ml