Thorax
ThoraxHR 2
nd
reconstr.
kV 130
Effective mAs 100
Slice collimation 1.0 mm
Slice width 5.0 mm 1.0 mm
Feed/Rotation 9.0 mm
Rotation time 0.8 sec.
Kernel B41s B70s
Increment 5.0 mm 0.7 mm
Image order cr-ca
CTDI
Vol
13.2 mGy
With studies of interstitial changes in the lungs, contrast
medium is not necessary.
MPRthick: ThoraxHR coronal
Image thickness 2
Distance between images 2
Number of images 70
For the 2
nd
reconstruction the Autoload into the MPR
Range on the 3D Card is activated. The images will
be automatically loaded into 3D, MPR, and a coronal
MPRthick Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.