143
Abdomen
AbdRoutine
kV 130
Effective mAs 95
Slice collimation 2.0 mm
Slice width 5.0 mm
Feed/Rotation 15.0 mm
Rotation time 0.6 sec.
Kernel B41s
Increment 5.0 mm
Image order cr-ca
CTDI
Vol
10.8 mGy
You could repeat the same protocol simply by clicking
the chronicle with the right mouse button for “repeat“.
E. g. when both non-contrast and contrast studies are
required.
Contrast medium IV injection
Start delay 50 60 sec.
Flow rate 4.0 ml/sec.
Total amount 100 ml