AngioRoutine 2
nd
reconstr.
kV 110
Effective mAs 90
Slice collimation 1.0 mm
Slice width 5.0 mm 1.25 mm
Feed/Rotation 9.0 mm
Rotation time 0.8 sec.
Kernel B31s B31s
Increment 5.0 mm 0.8 mm
Image order cr-ca
CTDI
Vol
8.2 mGy
Do not administer oral contrast medium, as this
impairs the editing of MIP/SSD/VRT images. Use water
instead if necessary.
* CARE Bolus may be used to optimize the bolus timing.
Set the ROI for monitoring scan in the abdominal
aorta at the beginning of the scan range with trigger-
ing threshold of 120 HU, or use manual triggering.
MIPthin: AngioRoutine sagittal
Image thickness 3
Distance between images 3
Number of images 60
For the 2
nd
reconstruction the Autoload into the
MIPthin Range on the 3D Card is activated. The images
will be automatically loaded into 3D, MIPthin, and
sagittal MIPthin Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.
Contrast medium IV injection
Start delay 20 25 sec.*
Flow rate 3.0 3.5 ml/sec.
Total amount 100 120 ml
161
Abdomen