169
Pelvis
Hip 2
nd
reconstr.
kV 130
Effective mAs 95
Slice collimation 1.0 mm
Slice width 5.0 mm 1.25 mm
Feed/Rotation 9.0 mm
Rotation time 1.0 sec.
Kernel B41s B70s
Increment 5.0 mm 0.8 mm
Image order cr-ca
CTDI
Vol
12.5 mGy
In case of 3D study only, mAs can be reduced up to
50%, images should be reconstructed with at least 50%
overlapping and kernel B10s.
If only one side is examined, it is advisable to enter the
side in the comment line on the routine card.
MPRthick: Hip coronal
Image thickness 3
Distance between images 3
Number of images 20
For the 2
nd
reconstruction the Autoload into the MPR
Range on the 3D Card is activated. The images will
be automatically loaded into 3D, MPR, and a coronal
MPRthick Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.

Table of Contents