171
Pelvis
PelvisRoutine
kV 130
Effective mAs 105
Slice collimation 2.0 mm
Slice width 5.0 mm
Feed/Rotation 18.0 mm
Rotation time 1.0 sec.
Kernel B41s
Increment 5.0 mm
Image order cr-ca
CTDI
Vol
11.9 mGy
* If the examination performed requires a full urinary
bladder, wait at least 3 minutes following IV admini-
stration of the contrast medium.
Contrast medium IV injection
Start delay 50 sec.*
Flow rate 2.0 – 3.0 ml/sec.
Total amount 100 – 120 ml