Page 128 (Model K)
Step 34 Reconnect the flow sensor and a new audible /
visual alarm Calibrate Flow Sensor should be
activated and replace the Connect Flow Sensor
alarm.
Recalibrate the flow sensor. Repeat Step 5 and then advance to Step 35.
Step 35 Press the Mode Select button to enter the Mode
select panel.
Step 36 Select Ventilation Off from the Mode Select panel. Press the confirm button.
Note: At this point the user can terminate the functional testing. If the user
would like to functional test all the functions of the ventilator they should
press the Mode Select button and select CMV from the Mode Select panel,
do not press the confirm button. Advance to Step 37.