(Model K) Page 129
18.1.16 Functional Test of CMV Mode
Step 37 For the functional test set the following parameters in the CMV preview mode:
BPM..................... 30
Ti.......................... 1 sec
PEEP ................... 5mbar
PIP....................... 20 mbar
FiO
2
..................... 21%
After pressing the confirm button the user will be presented with the
CMV screen.
Check that:
TTV ..................... is set to OFF. (This button will not appear
when functional testing with no flow sensor fitted).