HRX-OM-W058
Chapter 3 Transport and Setting Up
HRS-R Series 3.3 Installation
3-17
3.3.5 Wiring of operation signal output and alarm signal output
The operation signal output and alarm signal output are the outputs
generated by a contact signal to shown the status of the product.
The specifications of the contact for each signal output are as follows.
Table 3-3 Signal output contact spec. at the time of shipment
Contact output
Signal explanation
(Default setting)
Operation
Output current 1
(Terminal no.5,6)
Run status signal
A
At run
Contact closed
At stop
Contact open
With power supply
shut off
Contact open
Output current 2
(Terminal no.3,4)
Remote signal
A
At remote
Contact closed
At non remote
Contact open
With power supply
shut off
Contact open
Output current 3
(Terminal no.1,2)
Alarm signal
B
At generation
Contact open
Not generated
Contact closed
With power supply
shut off
Contact open
Tips
This product has three output signals which can be customized depending on
the customer’s application
Signals below can be output. Refer to the Communications Operation Manual
for more details.
Signal for completion of preparation (TEMP READY)
Signal for operation stop alarm
Signal for continuing operation alarm
Signal for selection alarm
Signal for operation start timer setting
Signal for operation stop timer setting
Signal for power recovery setting
Signal for freezing prevention setting
Contact input signal detection