MIC-10 / MIC-30 – USER MANUAL
Test range according to IEC 61557-2 for V
N
= 1000V: 1000k to 100.0G
Display range
for V
N
= 1000V
Note: For insulation resistance below R
ISOmin
there is no accuracy specified because the meter
works in the current limit mode in accordance with the following formula:
where:
R
ISOmin
- minimum insulation resistance measured without limiting the current
V
ISOnom
- nominal test voltage
I
ISOnom
- nominal current (1mA)
Measurement of leakage current
Calculated basing on resistance
measurements
I
Lmax
– maximum current at short circuit of leads,
resolution and units result from the measurement range of individual insulation resistance.
Measurement of capacitance
Measurement of capacitance is made only during R
ISO
measurement.
For measurement voltages below 100V and when measured resistance is below 10MΩ, the
measurement error is not specified.
Low-voltage continuity and resistance measurement
Measurement of continuity of protective conductors and equipotential bondings with 200 mA
current
Measuring range according to IEC 61557-4: 0.10 to 1999
Voltage at open terminals: <8V
Output current at R < 2: I > 200mA
Compensation of test leads resistance
Current flowing unidirectionally
Current flowing bidirectionally, average resistance is displayed