MIC-10 / MIC-30 – USER MANUAL
4 Memory of measurement result data
MIC-10 / MIC-30 Meters are equipped with a memory for storing test results (990 cells, each of
which may contain a set of measurements of R
ISO
and R
CONT
). The whole memory is divided into 10
memory banks with 99 cells in each bank. Thanks to dynamic memory allocation, each of the memory
cells can contain different quantity of single measurement results, depending on the needs. Optimal use
of the memory can be ensured in this way. Each measurement result can be stored in a memory cell
marked with a selected number and in a selected memory bank. Thanks to this, the user of the meter
can, at his/her option, assign memory cell numbers to individual measurement points and the memory
bank numbers to individual facilities. The user can also perform measurements in any sequence and
repeat them without losing other data.
Memory of measurement result data is preserved when the meter is switched off. Thanks to this,
the data can be later read or sent to a computer. The number of a current memory cell or memory bank
is not changed either.
Notes:
- Results of measurements performed for all measuring functions can be stored in one memory cell,
excluding R
X
and V .
- After entering the measurement result, the number of the cell is automatically incremented.
- It is recommended to delete the memory after reading the data or before performing a new series of
measurements that may be stored into the same memory cells as the previous ones.
4.1 Storing the measurement results in the memory
After completing measurement
press ENTER.
The cell is occupied
by the same type of
result, which is to be
entered.
Use and to preview the results.