EasyManua.ls Logo

Sony Trinitron PVM-14M4A - Secam Chroma; Secam R-Y Level; HP Width Adjustment

Sony Trinitron PVM-14M4A
92 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11-2. Writing HP.WIDTH (NORMAL) data
Note : Not applicable to groups of models 1, 2, 4, and 5.
1) Set the unit in the service mode.
2) Input 102 to HP.WIDTH (NOR).
11-3. HP POSITION adjustment
Note : 11-3. is not applicable to the group of models 3.
1) Input a SECAM color bar.
2) Set the HV-DL mode.
3) Set the unit in the service mode.
4) Adjust !HP POSmONI as shown in Fig.5-30.
Note: The same as 11-3. The phase relationship between the
beginning of IC317 @) PIN pulse and the input VIDEO
signal may be used for control.
/ H-SYNC
L
·····················••.•.••.•C• .. w.•.,.,.,.,.,.,.•.,.,. ,.__ V•SYNC
.... of plciun, -I
t
Bring the BLK edge to this
position by HP POSITION.
I-Beglaal"I) of plctu,e
I
Bring the BLK edge to this
position by HP WIDTH HN.
Fig. 5-30
11-4. HP WIDTH (H/V-DL) adjustment
Note : 11-4. is not applicable to the group of models 3.
1) Input a SECAM color bar.
2) Set the unit in the HV-DELAY mode.
3) Set the unit in the service mode.
4) Adjust !HPWID11i H/V-DELAYI as shown in Fig.5-30. (Note: Check
HP POSITION. If it is not in position, repeat 2) and 3).)
11-5. SECAM COL BALANCE
Note : 11-5. is not applicable to the group of models 3.
1) Input a SECAM color bar.
2) Connect the probe of an oscilloscope to TP306.
3) Set the unit in the service mode.
4) Adjust !SE.CAM COLOR BALANCER-YI so that the level in the ach-
romatic color will be straight.
5) Connect the probe of an oscilloscope to TP305.
6) Adjust !SECAMCOLORBALANCEB-YI so that the level in the ach-
romatic color will be straight.
11-6. SECAM CHROMA
1) Input a SECAM color bar.
2) Connect the probe of an oscilloscope to IC404 ® PIN orTP402.
3) Set the unit in the service mode.
4) Adjust !CHROMA SECAM! so that the peaks of waveforms will be
flush with each other as shown in Fig.5-31.
IC 404®PIN
Make Band D flush with each other.
A
D
--------------------
B
C
Fig. 5-31
11-7. SECAM R-Y LEVEL
l) Input a SECAM color bar.
20mVmax.
~
t
2) Connect the probe of an oscilloscope to IC404 @PIN orTP401.
3) Set the unit in the service mode.
4) Adjust IR-YLEVBL SECAMI so that the peaks of waveforms will be
flush with each other as shown in Fig.5-32.
IC404@PIN Make adjustment so that B=D.
60mVmax.
t
t
----------·-·
B
... --------
D
t
20mVmax.
Fig. 5-32
12. Writing adjustment results
Write adjustment results in the memory.
-37-

Table of Contents

Related product manuals