ST2683A Operation Manual Ver.1.3
16
ON
It is used for setting the PICK FUNC as ON.
OFF
It is used for setting the PICK FUNC as OFF.
Touch the corresponding key in soft key zone to select the relevant function.
NOTE: The following steps can be operated under the premise of setting the PICK FUNC as ON.
3.3.2 PICK ITEM
This function is used for selecting the pick item of the instrument.
Steps for PICK ITEM:
Touch the corresponding zone. The following items will be displayed in the soft key zone.
CURR
It is used for setting the PICK ITEM as CURR.
RESIS
It is used for setting the PICK ITEM as RESIS.
Touch the corresponding key in soft key zone to select the relevant function.
3.3.3 LOLI ONE
This function is used for setting the lower limit value of the instrument Bin One.
Steps for setting the LOLI ONE:
1. Touch this key to select this option and then the color of this option will become the color of
the cursor.
2. Touch this key (cursor) again to enter into the number soft keyboard, input the required lower
limit value (four-digit significant figures) and then press the ENTER key to confirm and return
to <PICKOUT SETUP> page.
The setting methods of UPLI ONE, LOLI TWO, UPLI TWO, LOLI STR and UPLI STR are the same as
LOLI ONE, the setting ranges for three groups of lower/upper limit value are:
1. When the PICK ITEM is set as CURR, the range is 1pA—1.25mA.
2. When the PICK ITEM is set as RESIS, the range is 100kΩ-10TΩ.
3. The lower limit value of each bin cannot be larger than the upper limit value, otherwise
“please re enter due to over-limit data” will be noticed.
4. The pickout ranges of three bins can be overlapped.
3.3.4 BEEP SET
This function is used for setting the pickout beep of the instrument.
Steps for BEEP SET:
Touch the corresponding zone. The following items will be displayed in the soft key zone.
BIN ONE
It is used for setting that the buzzer will alarm when the test value passes BIN ONE.