5.3 Measurement Accuracy
Measurement accuracy depends on measurement stability, temperature coefficient, linearity,
repeatability and correction interpolation error.
Checking the measurement accuracy should be taken under the following circumstances:
a Warm-up time should be more than 30 minutes.
b Correctly open the output terminal and perform zero clearing by pressing the
start key on the MEAS SETUP page.
5.3.1 Basic Accuracy for Resistance Measurement
SLOW, FAST: (This accuracy is obtained under the temperature of 20±5)
Test accuracy
Voltage ≥10V and I10nA :±2%
Voltage ≤10V or I ≤ 10nA :±5%