EasyManua.ls Logo

Sourcetronic ST2683A - Steps for bin Disp; Steps for PICK LIVA

Sourcetronic ST2683A
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ST2683A Operation Manual Ver.1.3
17
BIN TWO
It is used for setting that the buzzer will alarm when the test value passes BIN TWO.
BIN STREE
It is used for setting that the buzzer will alarm when the test value passes BIN STREE.
NG
It is used for setting that the buzzer will alarm when the test value of three bins fail.
OFF
It is used for setting the PICKOUT BEEP as OFF.
Touch the corresponding key in soft key zone to select the relevant function.
NOTE:
1. The BIN ONE enjoys the top priority in the three bins, followed by BIN TWO and BIN STREE.
BIN TWO and BIN STREE will not compare when BIN ONE passes, and so on, the test fails
when all three bins fail.
2. The pickout results will be displayed on the main display interface when the PICK FUNC is set
as ON. When one of the three bins passes, the main display interface will light up the green
light and indicates BIN1, BIN2 and BIN3. When all three bins fail, the main display interface
will light up the red light and indicates FAIL.
3.3.5 Steps for BIN Disp
Touch the corresponding zone. The following items will be displayed in the soft key zone.
ON
It is used for setting the BIN Disp as ON.
OFF
It is used for setting the BIN Disp as OFF.
Touch the corresponding key in soft key zone to select the relevant function.
3.3.6 Steps for PICK LIVA
Touch the corresponding zone. The following items will be displayed in the soft key zone.
ON
It is used for setting the PICK LIVA as ON.
OFF
It is used for setting the PICK LIVA as OFF.
Touch the corresponding key in soft key zone to select the relevant function.
NOTE:
1. When the PICK LIVA is set as OFF, if the PICK ITEM is set as RESIS, the upper limit value of
BIN1, BIN2 and BIN3 will be set as infinitely great and become unchangeable and drop out
of compare, namely, the pickout changes from interval pickout to single-point pickout; if the
PICK ITEM is set as CURR, the lower limit value of BIN1, BIN2 and BIN3 will be set as 0 and
become unchangeable and drop out of compare, namely, the pickout also changes from
interval pickout to single-point pickout.
2. When the PICK LIVA is set as ON, whether the PICK ITEM is set as RESIS or CURR, set the limit

Table of Contents