ST2683A Operation Manual Ver.1.3
30
5.1.6 Display Digits
Item
Resistance
4-digit significant figures
Leakage
current
Range 1correct to 100nA
Range 2correct to 10nA
Range 3correct to 1nA
Range 4correct to 100pA
Range 5correct to 10pA
Range 6correct to 1pA
Time
Correct to 0.1s
5.2 Test Signal
5.2.1 Output Voltage Range
Max. output DVC range of ST2683A test terminal: 1 1000V.
Max. output DVC range of ST2683B test terminal: 1 500V.
5.2.2 Output Voltage Accuracy
Voltage ≥ 10V: 1% ± 1V
Voltage < 10V: 10% ± 0.1V
5.2.3 Maximum Display Range
Parameter
Measurement Display Range
Resistance test
100.0kΩ ~ 10.00TΩ
Leakage current test
Range 1: 100uA ~ 1mA, input internal resistance 10kΩ
Range 2: 10uA ~ 100uA, input internal resistance 10kΩ
Range 3: 1uA ~ 10uA, input internal resistance 10kΩ
Range 4: 100nA ~ 1uA, input internal resistance 10kΩ
Range 5: 10nA ~ 100nA, input internal resistance 1MΩ
Range 6: 1nA ~ 10nA, input internal resistance 1MΩ