ST2683A Operation Manual Ver.1.3
29
5) Performance Index
5.1 Measurement Function
5.1.1 Measurement Parameters and Notations
RResistance
I: Leakage current
5.1.2 Range
Resistance Mode: AUTO, MANU (HOLD, UP, DOWN)
5.1.3 Trigger
Manual, external, bus
Manual: Press the TRIGGER” button on the panel, the instrument will test a DUT once and display
the result. This mode keeps in waiting mode when it is not used.
External: Test a DUT once and display the result when the instrument receives an external “start
up” signal from the footswitch of HANDLER interface on the rear panel.
BUS: The measurement of the instrument will be triggered through the communication interface.
5.1.4 Mode of Test Terminal
2-terminal measurement mode
HV (-): DCV output terminal INPUT: Sampling input terminal
5.1.5 Measurement Speed (when MEAS DISP & BIN DISP is set as OFF)
Speed
Parameter
FAST
SLOW
R
Meas/sec≤30ms
Meas/sec≤60ms
I
Meas/sec≤30ms
Meas/sec≤60ms