Setup
91369 Service Manual 2-34
Continuous Alarm Tones
The CONTINUOUS ALARM TONES key enables you to configure the monitor for continuous alarm tones. This
has no affect on the alarm period keys in the Tone Configuration menu.
Alarm Tone Access
The TONE ACCESS / ON/OFF key enables you to configure the monitor for user alarm tone access. The
TONE / ON/OFF key in the Alarm Tone, Remote Alarm, and Alarm Watch Tone menus is not displayed if
OFF is selected. This has no affect on the alarm period keys in the Tone Configuration menu.
Edit Drug List
The EDIT DRUG LIST key in the Biomed Level menu enables you to input and edit the master drug list from
any monitor. The master drug list can then be received by any monitor and stored in that monitor’s non-volatile
memory. The master drug list can store up to 16 drug names and associated data.
Prior to initially storing the drug list in a monitor (either via direct input or transfer), all monitors have blank drug
list names and display defaults that correspond to the standard drugs A through D (refer to Table 8). This set of
four default values is repeated for every fourth item in the drug list (for example, default settings for the 5th and
9th drugs in the list are identical to the settings for Drug A, the 1st drug in the list).
Touch EDIT DRUG LIST to display the Drug List Selection Menu (Figure 2-19). Selecting any drug name
from this list highlights that line. The drug list may be blank (if all the entries are blank) or display one or more
blank lines.
Table 8: Standard Drug Values
Key Neonatal Drug Adult Drug
A Dopamine or Dobutamine Dopamine or Dobutamine
B Isoproterenol Lidocaine, Pronestyl or Bretylium
C Tolazoline Nitroglycerine or Levophed
D Nitroprusside Nitroprusside