Chapter 6
Adjustment and Verification
P/N 79000035_SM, Version 1.5 6 - 13 Focus 30/35 Service Manual
2 Start Reticle
adjustment software
and level instrument.
A. Screw out the
instrument hight
adapter to position 3
of the scale and place
the instrument on the
collimator bench.
B. Set Collimator
wavelength to
position 1.
C. Select
Reticle
adjustment in Reticle
Maintenance &
Service.
D. Select check
box,Level the
instrument
E. Level instrument and
click Level OK
The Collimator
Monitor
Software starts
automatically.
Use the
software to view
the collimator
and instrument
reticle, set
Camera settings
to green light
(546nm).
Refer also to the
Software user guide
chapter in the Trimble
Collimator Bench
manual on:
Collimator Bench
manual
Illuminate the reticle
with the Illumination
tool (yellow LED).
Tolerance:
<10cc in both
directions.