Chapter 6
Adjustment and Verification
P/N 79000035_SM, Version 1.5 6 - 97 Focus 30/35 Service Manual
Tracker test
Performance
Fig. 6-80 Tracker receiver focus
3 Exit software Push
Exit button and
perform Tracker test.
Item Performance Action Result Notes
1 Start up instrument /
PASS.
See page 7-3 how to set-
up the instrument.
2 Start the test. A. Aim to a precise
indoor prism in 15m
to measure the
Tracker signal
performance.
B. Select
Tracker Test
from Tracker
adjustment
in
Maintenance &
Service, and start the
test measurement.
Tolerance:
15m ±5cm
Tolerance:
Pixel Qty.
> 1250
Luminance
244
Prism 62mm SECO
Tolerance:
Pixel Quantity and
Luminance will be
shown.
Item Performance Action Result Notes