Chapter 6
Adjustment and Verification
Focus 30/35 Service Manual 6 - 64 P/N 79000035_SM, Version 1.5
Reference fibre adjustment
Performance
4 Adjust Receiver fibre
position.
A. Turn the two position
screws counter-
clockwise, until the
fibre is unaligned.
B. Turn the two position
screws clockwise
until the fibre is
centered to the cross-
hair.
C. Knock slightly on the
ferule holder to
remove tensions.
D. Verify alignment and
lock screws.
Tolerance:
<10“ = 1/3 of a
garaduation on
the Collimators
cross-hairs.
1.3 mm ball Allan
key, see Fig. 6-48 on
page 6-63 .
The CM software
shows the Delta X/Y
values in green if it’s
in tolerance.
Glue MS 1855 on red
points.
Fig. 6-50 Receiver fibre alignment
5 Check Receiver fibre
adjustment again and
exit program.
A. Repeat item
2 to 4
check the correct
focus and position
again.
B. Push
Exit to leave the
program.
Tolerance:
<10“ = 1/3 of a
graduation on
the Collimators
cross-hairs.
The CM software
shows the Delta X/Y
values in green if it’s
in tolerance.
Item Performance Description Result Notes
1 Prepare instrument. Mount the instrument on
the Collimator bench or
on a stable base.
Item Performance Action Result Notes