Chapter 6
Adjustment and Verification
Focus 30/35 Service Manual 6 - 16 P/N 79000035_SM, Version 1.5
Fig. 6-9 Check HA / VA collimation - CHz / CV values in tolerance
6 Check HA/VA
collimation errors -
CHz correction value
is out of tolerance, CV
is in tolerance.
A message box appears:
CHz is out of tolerance and
needs to be adjusted in
Adjust HA colllimation
.
Condition:
CHz: >30” /
0,0083gon
CV: 30” /
0,0083gon
Quit the message box with
OK and go to see
“Adjustment of Reticle
F30 in Hz direction
keeping the vertical
orientation” on page 6-23.
No value will be
stored in to the MCB
board.
Item Performance Description Result Notes