Chapter 6
Adjustment and Verification
Focus 30/35 Service Manual 6 - 32 P/N 79000035_SM, Version 1.5
4 Adjust Hz collimation. A. The instrument
moves automatically
to the adjustment
position AP Hz.
Click Holding mode
ON.
B. Adjust the
instruments reticle in
coincidens to the
collimators reticle.
Keep attention to the
erecting of reticle see
Fig. 6-15 on page 6-
22 .
C. The vertical
adjustment screws
should hold the
reticle in the
displayed vertical
reference position.
D. Repeat item
3 and 4
until the tolerance is
reached.
Tolerance:
CHz <10cc
CV <30cc
Tolerance:
CHz <10cc
CV <30cc
To shift the reticle
turn the opposite
adjustment screws in
the same direction
with 1,5 allan key.
Item Performance Action Result Notes