Chapter 6
Adjustment and Verification
Focus 30/35 Service Manual 6 - 44 P/N 79000035_SM, Version 1.5
Fig. 6-31 Store new VA collimation after referencing with tool
6 Adjust Hz collimation. A. The instrument
moves automatically
to the adjustment
position AP Hz.
Click Holding mode
ON.
B. Adjust the
instruments reticle in
coincidens to the
collimators reticle.
Keep attention to the
erection of reticle see
Fig. 6-15 on page 6-
22 .
C. The vertical
adjustment should
hold the reticle in the
displayed vertical
reference position.
Tolerance:
CHz < 10cc
CV < 30cc
AP Hz = Hz+CHz
To shift the reticle
turn the opposite
adjustment screws in
the same direction
with 1,5 allan key.