Chapter 6
Adjustment and Verification
Focus 30/35 Service Manual 6 - 84 P/N 79000035_SM, Version 1.5
8 Complete EDM Offset
determination.
After completing the
baseline measurement for
EDM offset
determination, the new
value will be stored
temporary in the EDM.
The old and new
EDM Offset is
displayed.
At least 4x prism
targets are required.
9 Complete validation
and store EDM Offset
value.
A. Click
Measure to start
the distance
measurement for
validation
B. Click
Position to
move instrument to
the next target.
C. After successful
validation, the new
EDM offset value
can be stored
confirming with
Yes,
otherwise the old
value will be restored
in the EDM.
The measure
result and
standard
deviation is
shown after
each
measurement.
At least 1x prism
target and 2x DR
targets are required.
Fig. 6-70 Complete the EDM offset determination
10 Exit software. Click
Exit button.
Item Performance Action Result Notes