Chapter 6
Adjustment and Verification
Focus 30/35 Service Manual 6 - 144 P/N 79000035_SM, Version 1.5
Trouble shooting: A. The status lamps for
IM, Hz, V and Drv
are not shown in
green.
B. Save measure values
and log file. Push
button
Store Hz/V
measure data
.
C. The test shows some
calibration values
marked with red.
Check connection between PC and
instrument, the driver or re-boot the
instrument.
The files will be stored on hard drive and
can be attached in Remedy for
evaluation.
The calibration was not successful.
Check the tab pages Excentricity,
Amplitude and Ambiguity. The original
calibration values could be restored
separated by axis pushing
Restore Hz or
Restore V.
5 Check diagrams on tab
pages Hz-/V-
Excentricity:
Press tab page
Hz-Exc. or
V-Exc., see Fig. 6-112 on
page 6-145 .
The diagrams
should be
between the red
min./max lines.
Green spikes
means sensor
CIS1 can’t read
a value.
Black spikes
means sensor
CIS0 can’t read
a value.
-Corresponding
spikes in green
and black
shifted 200gon.
Select CIS1 or CIS0
to separate the
diagrams.
-Clean CIS1 sensor.
The position is shown
moving the cursor
over the sensor name.
-Clean CIS0 sensor.
-Clean the encoder
disk.
Item Performance Action Result Notes