9. Use a dry, grease-free cloth to
clean the surface of the spark
stand body.
10. Use a dry, grease-free cloth or a
bottle brush with plastic bristles
to clean the light channel (1).
Fig. 24
11. Install the spark stand insert (1).
Use two fitted pins (3) in
order to center the spark
stand insert.
Ensure that the spark stand
insert is flush with the top
edge (2) of the spark stand.
Fig. 25
12. Install the spark stand plate (1).
13. Insert the electrode (2) into the
spark stand opening.
Fig. 26