XRV320 and XRV350 Overview : XRV320 and XRV350 (Front Panels)
The cathode (negative) generator front panel of the
XRV320 and XRV350 models use the same LED
Indicators as the unipolar XRV160 showing the status
of the Generator. Green Indicators signify normal
operational mode while red indicators signify a fault
condition. A fault or red indication prevents high
voltage operation. Fault conditions are “latched”
meaning that a reset must be employed before re-
starting the generator. Once the fault clears, the red
indicator will extinguish. The anode (positive)
generator does not have any indicators since all of
the monitoring and fault status indicators originate
from the control board within the cathode (negative)
generator. Additionally, fault status and ready conditions are monitored from the I/O connector on the
rear panel and the GUI Interface software. A removable air filter keeps dust and dirt from entering the
chassis.