XRV Setup and Installation:
There are several installation configurations available for the XRV Subsystem. The most economical
system employs the use of a single ended X-ray tube, the XRV Generator, Personal PC, X-ray Cable, Input
Power / Communication Cables and Water Cooler/ Chiller. The XRV Generator includes ready to install
software that can be loaded into the user’s personal PC. A minimal setup for a standard Unipolar
subsystem is illustrated on the following page. The user must provide a personal PC for installing the
included software or elect to interface their own controls via the I/O from J2 (25 Pin subminiature D
connector) on the rear of the XRV Generator. Information regarding the I/O connector will be covered
later in this manual. It is imperative the installer follow all safety regulations and codes for installing this
equipment. Details such as input power requirements, wire sizing, and physical placement of each
component is critical for safe and trouble free operation. Do not install power wires or high voltage
cables in high traffic areas where they could be damaged by heavy equipment. Wire lengths should be
considered for voltage drops and thermal management. The X-ray tube