ersion N.NN
The display will show the
software version …..
N.NN = software version
Calibrate Mode
Please wait …
….. then change to show the
current status.
Release the Set Temperature
and Set Timer Buttons.
Release
The Sample Cup will start to
heat up to 55°C as measured
by the Setaflash Series 3
Flash Point Tester.
When the instrument measures
the Sample Cup temperature as
55°C the display will countdown
from 300 seconds while the
temperature stabilises.
Calibrate Mode
Please wait 300
°C 55°C
Issue K
10 Jun 2007
23