Pubn No: 30000H0K
5.4. CALIBRATION CHECK
Switch the instrument OFF at the
ON/OFF Power Switch on the
rear of the instrument. The
calibration is stored.
Press hold down the Set
Temperature
and Set Timer
Buttons and …..
….. switch the Setaflash Series
3 ON at the ON/OFF Power
Switch on the rear of the
instrument.
Calibrate Mode
Please wait …
V
ersion N.NN
The display will show the
software version …..
N.NN = software version
….. then change to show the
current status.
Press & Hold Down
Switch OFF
Switch ON
Issue K
10 Jun 2007