Pubn No: 30000H0K
Release
Release the Set Temperature
and Set Timer Buttons.
°C 55°C
The Sample Cup will start to
heat up to 55°C as measured
by the Setaflash Series 3
Flash Point Tester.
When the instrument measures
the Sample Cup temperature as
55°C the display will countdown
from 300 seconds while the
temperature stabilises.
Calibrate Mode
Please wait 300
When the countdown reaches 0
a warning beep will sound …..
….. and the display will change
to request the temperature as
shown on the thermometer.
Thermometer T? >
55.0°C
Issue K
10 Jun 2007
26