4.4. RUNNING A TEST
When the Sample Cup has
stabilised at the set temperature
a warning beep will sound …..
XX.X
o
C TT mins
Ready
XX.X
o
C AUTO Yml
Ready
Note: Sample Size:
Ambient to 100°C = 2ml.
Above 100°C = 4ml
FAME products = 2ml only
….. and the display will show
“Ready”.
Press & Release
X
X.
X
o
C
TT seconds
WARNING: DO NOT INJECT
SAMPLE WITH THE PILOT OR
TEST FLAME LIT.
The display will show the
countdown.
Press then release the Set Timer
Button
to start the timer
countdown.
Load the Syringe with sample
and inject the sample into the
Sample Cup through the Filler
Orifice. Remove the Syringe.