ERROR MESSAGE AND DESCRIPTION RECOMMENDED ACTION
Remove BI from Well <#>. Scan BI First.
The BI was not scanned before inserting into well.
Scan BI fi rst and then place into any empty well. If the fault returns, contact STERIS.
Invalid Barcode, Scan Again.
The barcode scanned was not recognized as a
valid barcode.
Check BI type and scan BI again. If the fault continues, contact STERIS.
Expired BI, Please Discard.
The BI is expired or the Date setting is not correct.
Check expiration date of BI, check date setting, and scan BI again. If the fault
continues, contact STERIS.
BI Not Activated, Well <#>.
The corresponding well has started a reading but
either there was no BI placed into the well or the
BI was not activated.
Verify that a BI is in the well and that it is activated. Press ENTER to acknowledge the
message. If this error continues to occur following verifi cation that a BI is present and
activated, contact STERIS or an authorized dealer.
Instrument Failure. <error code>.
Irrecoverable diagnostic fault.
Contact STERIS or an authorized dealer.
The Incubator is unusable. Discard a ected sample(s).
Return BI To Well <#>.
The BI was removed from the well before the
incubation period was over.
Return BI to well <#> before timer reaches zero
BI Prematurely Removed.
The BI was removed from the well for longer
than 10s.
Discard sample as per BI instructions. Press ENTER to acknowledge the message.
Remove All BIs.
The BI is in the well during warm-up and
diagnostics test.
Discard sample as per BI instructions. Press ENTER to acknowledge the message.
Power Lost, Shutting Down . . .
Device is unplugged or lost power.
Return power to the device before timer reaches zero. Check power is plugged into
the wall and power cord inlet.
Discard BIs.
Power was interrupted for longer than 60 seconds.
Discard sample as per BI instructions. Press ENTER to acknowledge the message.
No message, but error code next to well status.
Step 1: Error message may have been previously acknowledged. Remove BI(s) from
well(s) with error code and discard BI. If error continues, go to step 2.
Step 2: Clean wells per the Operators Manual. Cycle power to the device by
unplugging for > 2 minutes. If fault returns, contact STERIS or an authorized dealer.
Note: Do not attempt while actively incubating BIs in another well.