3
en
Celerity Incubator Figure Key . . . . . . . . . . . .1
Safety Information. . . . . . . . . . . . . . . . . . . . .3
A. Product Description. . . . . . . . . . . . . . . . .3
B. Intended Use . . . . . . . . . . . . . . . . . . . . . .3
C. Specifications . . . . . . . . . . . . . . . . . . . . .4
D. Getting Started . . . . . . . . . . . . . . . . . . . .5
E. Routine Operations . . . . . . . . . . . . . . . . .6
WARNING:
For indoor use only.
Do not immerse.
Clean only with recommended products.
Unit does not contain any user serviceable parts. Return to STERIS for authorized repairs (Call
authorized sales representative or STERIS Customer service).
To avoid personal injury or damage to device use the properly rated power cord supplied.
The Incubator is intended to be used solely with Celerity BIs. Do not substitute any other products for
the Celerity BIs.
Do not remove BI from well during incubation for more than 10 seconds until result is shown. Removal
of BI may invalidate test.
Caution:
To avoid possible environmental contamination, dispose of the Incubator per applicable government
regulations.
To avoid incorrect readings, place the Celerity BI vial so that it does not come into direct contact with
chemical indicators or tape.
Do not touch the hot surface of the metal incubator block.
Do not use abrasive materials to clean or disinfect the Incubator.
To avoid incorrect readings, do not place the Incubator in an environment exposed to direct sunlight or
strong incandescent light.
To avoid incorrect readings, do not move the Incubator while incubation is in-progress.
Do not spray water, cleanser, or disinfectant on or into the Incubator.
Well blocks may be HOT. Use caution when inserting BI.
A. Product Description:
The Celerity Incubator is designed exclusively to incubate and read activated Celerity Biological Indicators (BIs).
The device serves as both an Incubator and a fl uorescence detector. The Incubator heats the activated BIs to
temperatures which promote the metabolism and subsequent detection of viable test organisms under evaluation.
While the activated BI is incubating, viability is determined by an increase in a fl uorescent signal from the enzymatic
breakdown of a fl uorogenic substrate present in the media.
B. Intended Use:
Use the Celerity Incubator to incubate and automatically read STERIS Celerity Biological Indicators for Steam and
Vaporized Hydrogen Peroxide sterilization at 55°C-60°C for a fl uorescent result.
Refer to the Instructions for Use for the Celerity BIs to confi rm the validated readout times.
Table of Contents
F. Options. . . . . . . . . . . . . . . . . . . . . . . . . . .7
G. Cleaning and Care of the Incubator . . . . 8
H. Calibration and Diagnostics . . . . . . . . . .9
I. Troubleshooting. . . . . . . . . . . . . . . .9 & 10
J. Warranty. . . . . . . . . . . . . . . . . . . . . . . . .12
K. Disposal . . . . . . . . . . . . . . . . . . . . . . . . .12
L. Contact Information. . . . . . . . . . . . . . . . 12

Table of Contents