10
en
In the case of a temporary user error, the Incubator will sound an alarm (one
beep every fi ve minutes) and the LED adjacent to the well will fl ash a yellow/
amber light. The LCD screen will display an error message at the top the screen
and an “ERR“ next to the well number. The operator must take corrective action
based on error presented, then press the ENTER button to acknowledge the
alert and silence the alarm.
See Troubleshooting Guide below for sample error messages.
1
2
3 ERR 13
4
5
6
7
8
BI Prematurely Removed
In the case of a permanent error, the LED adjacent to the well will turn to a
solid yellow/amber light. The LCD screen will display an “ERR“ next to the well
number but no error message at the top since no user action can correct the
issue. These are permanent errors that may require the user to contact STERIS
or an authorized dealer. Do not use the a ected wells. The operator must press
the ENTER button to acknowledge the alert.
1
2
3
4
5
6
7
8 ERR 01
10. Allow the BI to equilibrate for a minimum of 10 minutes.
11. Verify the temperature reading on the thermocouple thermometer. If the temperature falls out of range for
incubation, contact STERIS or any authorized dealer. Do not use failing wells. The other panel of wells may
still be operational.
12. If the recorded temperature is within the range for incubation, abort the reading by removing the BI. Dispose
of the BI as per BI instructions.
I. Troubleshooting Guide