4
en
C. Specifi cations:
AC Input Voltage: 100-240 VAC +/-10%
AC Input Frequency: 50-60 Hz
Input Current: 1.0A (rms) max
Overvoltage: Category II
Operating voltage: 12 VDC
Operating current: 3.0 A max
Well Temperature
Specifi cations:
55-60°C
Detection Modes: Fluorescence
Emissions Wavelength: 410-450 nm
User Interface: LCD and Navigation Buttons
Interface with Optional
Printer:
RS-232 port
Audible Alarm:
Low: 50-60 dB
Medium: 70-80 dB
High: 80-100 dB
Weight: 3 lbs 10 oz (1.64 kg)
Dimensions:
6.5 (W) x 6.0 (H) x 11 (D) inches
165 (W) x 153 (H) x 280 (D) mm
Incubator Environmental Conditions
Temperature: 15°C-40°C/59°F-104°F
Relative Humidity: 20-80%
Altitude: Up to 2000 m
Pollution degree: 2
Degree of ingress
protection per IEC 60529:
IPX0
Caution:
To avoid incorrect readings, do not place the Incubator in an environment exposed to
direct sunlight or strong incandescent light.
Device Safety Information
The Celerity Incubator meets the applicable requirements for the following standards:
• Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use Part 1: General
Requirements [UL 61010-1:2012 Ed.3+R:19Jul 2019]
• Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use Part 1: General
Requirements [CSA C22.2#61010-1-12: 2012 Ed.3+U1; U2; A1
• Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use Part 2-010: Particular
Requirements for Laboratory Equipment for The Heating of Materials [UL 61010-2-010:2019 Ed.4; CSA
C22.2#61010-2-010:2019 Ed.4]
• EMC Directive 2014/30/EU
• EN 61326-1; EN 55011
• FCC CFR 47, Part 15, subpart B:2007
• ICES-003, Issue 7: 2020
• AS/NZS CISPR 11:2011
• IEC 61326-1
• EMC Testing CISPR 11
• KS C 9811
Symbol Defi nition Symbol Defi nition
Manufacturer Tested to comply with FCC standards
Mark of Conformity to European Directives
Compliant to all applicable ACMA regulatory
arrangements (RCM)
ETL Listed to US and Canadian Safety Standards Catalog number
NOTE: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to
part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference
when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate
radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful
interference in which case the user will be required to correct the interference at the user's own expense.
Product & Packaging Label Symbols