19 620 02 0117
55
Operation
CAUTION!
Malfunction of the device or its components
Various injuries are possible!
Immediately comply with all status and error messages which
appear during the treatment (see Operating Manual of the
handpiece).
CAUTION!
Over extended periods, the noise of the pulses can be per-
ceived as unpleasant!
Offer ear protection to the patient.
Recommendation: The user should also wear ear protection.
CAUTION!
Danger of injury from inadvertent pulse triggering
Trigger pulses only when the handpiece is in contact with the in-
tended treatment zone.
Avoid applying excessive contact pressure with the transmitter to the area to be
treated. Excessive pressure is not needed for the success of the treatment.
CAUTION!
The pulse transmitter surface will become hot!
Extended skin contact can lead to minor burns!
Interrupt treatment after a maximum of 6,000 pulses.
Do not apply more than 300-400 pulses to the same spot during treatment.
Place the handpiece back in the handpiece holder after the treatment.