19 620 02 0117
65
TechnicalSpecications
7 TechnicalSpecications
7.1 TechnicalSpecications
D-ACTOR
®
200
D-ACTOR operating mode
single pulse mode,
continuous pulse mode
HP 21700.0001: 1-21 Hz / 1-5 bar in steps of 0.1 bar
HP 21700.1001: 1-21 Hz / 0.3-5 bar
eff
operating mode V-ACTOR:
- standard
- when using an
external display
31 Hz / 2,4 bar
1 - 35 Hz / 1 - 5 bar
operating mode VACU-
ACTOR:
intensity
with foot switch: // with tablet:
3 - all values adjustabler
according Tabelle 4 -3
pressure VACU-ACTOR max 200 mbar
under pressure VACU-ACTOR min. - 500 mbar
Mains input voltage 220 - 230 VAC // 115 VAC
Mains frequency 50 - 60 Hz // 60 HZ
Mains fuse T8AL / 250 VAC
Power consumption max. 500 VA
Compressed air supply internal
Compressed air output 1 – 5 bar
Ambient temperature
during operation
10° – 40°C
Ambient temperature
during storage and transport
0° – 60°C frost free
Ambient pressure
during operation
800 – 1060 hPa
Ambient pressure
during storage and transport
500 - 1060 hPa
Air humidity
during operation
5 – 55%, non-condensing
Air humidity
during storage and transport
5 – 95%, non-condensing
Control device weight 25 kg
Housing dimensions (W x H x D) 487 x 191 x 465 mm
ClassicationaccordingtoMDD Class IIa device
Protection against the ingress
of water
IPX1