Electromagnetic compatibility
Instructions for use • VITOMEAGLE • TVA900_EN_V1.0_05-2023_IFU_CE-MDR 32
10 Electromagnetic compatibility
The product is suitable for use in professional healthcare settings. Professional healthcare facilities
include physician offices, dental offices, limited care facilities, freestanding surgical centers,
freestanding birth centers, multiple treatment facilities, hospitals (emergency rooms, patient rooms,
intensive care, surgical rooms, outside the HF-shielded room of an ME system for MRT).
The described product has been tested as a system with the following devices. Relevant
electromagnetic compatibility (EMC) information can be found in the "Electromagnetic compatibility
(EMC)" section of the instructions for use for the devices.
Device Item number
ARTip CRUISE UR500B
ARTip CRUISE UR500C
ARTipCRUISE Adaptor for VITOMEAGLE UR007
IMAGE1 S CONNECT II TC201
IMAGE1 S 4U-LINK TC304
Power LED RUBINA, OPAL1 NIR/ICG TL400
32" 4K/3D Monitor TM350
The EMC warning statements, precautions, notes and Emission/Immunity limits specified in the
instructions for use for the basic devices also apply to the applied part described in these instructions
for use.
The emission characteristics of this product make it suitable for use in industrial areas as well as
in hospitals (CISPR 11 Class A) and other professional healthcare environments. If it is used in a
residential environment (for which CISPR 11 Class B is normally required), the product may not
offer sufficient protection for radio transmission operation. The user might need to take
mitigation measures, such as relocating or re-orienting the product.