Maintenance Manual
10
2.2 LUBRICATION REQUIREMENTS
The only components of the bed needing periodic verification and lubrication are the four
actuator screws and the clevis pin holding the head actuator to the head section lever. They
should be checked every year and lubricated every two years.
The bed uses oil-impregnated shoulder spacers at hinge points. Do not lubricate these shoulder
spacers. When shoulder spacers are found worn, replace them.
ACTUATOR SCREW LUBRICATION
Required Tools:
No. 2 Phillips Screwdriver 1/4" Ratchet (w/6" extension) and 5/16" socket
OG2 Grease Brush Bung ee Cord
Procedure:
Clevis Pin
1. Raise the bed fully up and apply the brakes.
2. Remove if applicable the head section plastic cover (optional).
3. Fully raise the head section.
4. Remove the head board.
5. Unplug the bed power cord from the wall outlet.
6. Apply grease all over the clevis pin.
7. Fully raise and lower the head section several times to spread the grease.
C
C
A
A
U
U
T
T
I
I
O
O
N
N
A
Figure 2.2
B
B

Table of Contents

Other manuals for Stryker FL28C

Related product manuals