Introduction Chapter 1
5
1. INTRODUCTION
This manual is designed to assist in the servicing of the Stryker's FL28C Med/Surg Beds. Read
it thoroughly before beginning any service on the bed. Qualified maintenance personnel should
be able to refer to this manual at all time when servicing the bed.
This Maintenance Manual is an integral part of the bed and should be included if the bed is sold
or transferred.
1.1 SPECIFICATIONS *
Safe Working Load** 500 lb (227 kg)
Scale System (Optional)
- Capacity
- Accuracy
- Operating Angular Range
Patients weighing up to 500 lb (227 kg)
± 2 % for weight from 100 to 500 lb (45.3 to 227 kg)
± 2 lb for weight under 100 lb (45.3 kg)
-12° to +12°
Overall Length/Width
- Siderails Up
- Siderails Down
94 5/8 x 40" (240.34 cm x 101.6 cm)
94 5/8 x 39" (240.34 cm x 99.06 cm)
Weight w/Boards 472 lb (214.1 kg)
Patient Sleep Surface 35 x 80" (89 x 203 cm) adjustable to 82" (208 cm) and
84" (213 cm)
Recommended Mattress Size
Mattress Maximum Thickness
35 x 80" (89 x 203 cm); 35 x 82" (89 x 208 cm); 35 x
84" (89 x 213 cm)
6" (15.24 cm)
Min/Max Bed Height 14 1/2 to 29" (36.8 to 73.7 cm)
Fowler Angle 0 to 61°
Knee Gatch Angle
- with Auto Contour (optional)
- without Auto Contour
0 to 24°
0 to 32°
Trendelenburg/Reverse Trendelenburg +14 to -14°
Environmental Conditions
- Transport and Storage
- Ambient Temperature
- Relative Humidity
- Atmospheric Pressure
- Operating ***
- Ambient Temperature
- Relative Humidity
- Atmospheric Pressure
-40 to 70°C (-40 to 158°F)
10 to 100%
500 to 1060 hPa
18.3 to 26.7°C (65 to 80°F)
20 to 80% without condensation
700 to 1060 hPa
Electrical Requirements**** - all electrical
requirements meet CSA C22.2 No. 601.1,
UL 60601-1 and IEC 60601-1, 60601-2-38
specifications.
100V, 50-60Hz, 7.5A - Two 250V, 10A Fuses
120V, 50-60Hz, 4.0A (9.8A w/120V Optional Auxiliary
Outlet) - Two 250V, 10A Fuses
200V, 50-60Hz, 3.2A -Two 250V, 6.3A Fuses
220V, 50-60Hz, 2.9A -Two 250V, 6.3A Fuses
240V, 50-60Hz, 2.7A -Two 250V, 6.3A Fuses
* Stryker pays special attention to product improvement and reserves the right to change specifications
without notice.
** The Safe Working Load specified is the sum of the mattress and accessory weight (100 lb/45.4 kg) and the
patient's weight.
*** Operating environment recommended to ensure the scale system (optional) precision.
**** The device has a 10% duty cycle.