Mai ntenance Procedures Chapter 4
19
7. Reverse the above steps to install the new siderail assembly.
8. Test the siderail movement and all controls of both control panels, including the nurse call
optional) and the communications package (optional), for proper operation before returning
the bed to service.
HEAD RAIL REPLACEMENT
Please contact the Technical Service (see section 1.2) for information regarding the
replacement of a head end rail.
HEAD SIDERAIL MECHANISM REPLACEMENT
Please contact the Technical Service (see section 1.2) for information regarding the
replacement of a head end siderail mechanism.
FOOT SIDERAIL ASSEMBLY REPLACEMENT
Required Tools:
1/2" Socket Wrench 1/2" Wrench Bungee Cord
Procedure:
1. Run the bed fully up and apply the brakes. Raise the siderail needing repair.
2. Run the Knee Gatch fully up and unplug the power cord from the wall receptacle.
3. Remove if applicable the foot section plastic cover (optional). Lift and fold the foot section
back toward the head end of the bed. Secure its position using a bungee cord.
4. Using a 1/2" socket wrench and a 1/2" wrench, remove the four locknuts/bolts (A) holding
the siderail assembly to the support and remove the assembly. Support the siderail
assembly while removing the last bolts.
5. Reverse the above steps to install the new siderail assembly.
6. Test the siderail for proper operation before returning the bed to service.