Maintenance Procedures Chapter 4
43
9. As indicated, do not touch the bed until the end of this step. Once the last step is over, the
system will automatically bring in the main maintenance menu.
10. To get back to the normal scale display, press ZERO.
11. Before returning the bed to service, Test the scale accuracy by following the operations
described above under "Verifying Scale Accuracy".
N
N
O
O
T
T
E
E
It is possible to leave the calibration process any time before step 3 is completed by pressing the
ZERO key. The preceding calibration data will be restored.
If the bed is placed at the wrong angle in step two or three, the same message will reappear on
the display until the bed is placed at the angle requested on the display.