Chapter 6 Operation
www.supmea.com
-35-
6.5. Configuration details
NO.
Parameter
Setting
mode
Password
level
Parameter
range
Default
1-Flow
1-0
Flow range
Figure
User
0-99999
35.000
Set the maximum flow limit value. Used to calculate the frequency,
output current limit calculation and alarm threshold calculation, etc
1-1
Flow unit
Option
User
L
m
3
Kg
t/s
min
h
gal/m
gal/h
m
3
/h
Choose volume unit ,such as L, m
3
, gal; the density will not calculated;
Choose mass unit such as Kg, t; need 1-2 density parameter.
1-2
Fluid density
Figure
User
0.000-99.000
1.000
Used to calculate the mass flow, Q
M
=ρV
M
. When the flow unit is
volume, this parameter are not displayed. Density unit
:
g/cm
3
1-3
Time
constant
Figure
User
0-99S
2s
Damping coefficient of the filter, select the average selected within the
time parameter as the real-time flow.
1-4
Flow
resection
Figure
User
0-10%
1%
NO.
Parameter
Setting mode
Parameter
range
Default
1
Sensor drift diameter
Option
3-2000
50
2
Flow range
Figure
0-99999
35.000
3
Sensor coefficient
Figure
0-99999
1.000
4
Zero correlation
Figure
0-99999
0.0
5
Accumulation reset
Option
YN
N
6
Flow remove
Figure
0-99%
1%
7
Time constant
Figure
0-99S
2s